Product specification
Product model: Intelligent Disintegration Time Limit Tester BJ-2 (Intelligent Double Cup)
product manual:
●The disintegration time tester is used to check the disintegration of solid preparations under specified conditions.
Applicable standards:
● "2010 Pharmacopoeia of the People's Republic of China" edition.
●Enterprise standard "Disintegration Time Limit Tester" Q/12XQ0188-2010.
main feature:
●Double cups, twelve tubes.
●Double cups and two channels, each channel runs independently and can be controlled separately; automatic timing shutdown.
●Automatic temperature control, high temperature control accuracy.
●Imported stainless steel is used for hanging blue and boom.
●Adopt micro water pump circulating water flow uniform heat system, water bath temperature is uniform.
●Fully automatic intelligent control of three parameters: temperature, time and round-trip frequency.
● Time parameters can be preset at will; preset values and real-time values are displayed in time-sharing.
●Automation, automatic detection, automatic diagnosis, automatic alarm.
Technical index:
●Number of lifting baskets 2
●The number of glass tubes in the gondola 12
●Flat bottom beaker volume 1000ml
●Return frequency of gondola (30~32) times/min
●Return stroke of gondola (55±1)mm
●Minimum spacing of screen mesh (25±2) mm
● Timing range (1~900) minutes
● Timing accuracy ±0.5min
●Temperature preset range (20~45)℃
●Automatic temperature control accuracy ±0.5℃
●Power of heater 500W
●Continuous working time more than 24 hours
●Power supply 220V/50Hz
●Power 600W
●Overall dimensions (40×30×42) cm3
●Weight 16kg